User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Honeywell 8716:Digital Indicator RH DATE: 28-Sep-94 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 13 NUMBER OF LINES: 105 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # This procedure is run with the MET/CAL Configuration #1 loaded. 1.001 ASK- R P F W 1.002 ASK+ X 1.003 MMFC C1 1.004 HEAD ÍÍ INITIAL CONDITION ÍÍ 1.005 DISP ÿ Connect the 5700A and the UUT as follows: 1.005 DISP ÿ [27][91]1m 5700A TO 8716 1.005 DISP ÿ ACCROSS THE CURRENT RESISTOR 1.005 DISP ÿ OUTPUT V ê A HI ÄÄÄÄÄÄÄÄÄ INPUT 9 rear terminal 1.005 DISP ÿ OUTPUT V ê A LO ÄÄÄÄÄÄÄÄÄ INPUT 10 rear terminal 1.006 HEAD {} 1.007 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.008 HEAD {º FULL SCALE AND LINEARITY TEST º} 1.009 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.010 HEAD {}ÍÍ FULL SCALE AND LINEARITY TEST ÍÍ 1.011 JMP 2.001 1.012 EVAL 2.001 5700 4mA S 2W 2.002 MEMI Use keyboard to enter actual UUT reading in Percent. 2.003 ACC 0.0pctrh 0.0003U 2.004 MEME 2.005 MEMC pctrh 0.1U 4mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 3.001 5700 5.6mA S 2W 3.002 MEMI Use keyboard to enter actual UUT reading in Percent. 3.003 ACC 10.0pctrh 0.0003U 3.004 MEME 3.005 MEMC pctrh 0.15U 5.6mA #! Test Tol 0.15, Sys Tol 0.0003, TUR 500.000 (>= 4.00). 4.001 5700 7.2mA S 2W 4.002 MEMI Use keyboard to enter actual UUT reading in Percent. 4.003 ACC 20.0pctrh 0.0003U 4.004 MEME 4.005 MEMC pctrh 0.2U 7.2mA #! Test Tol 0.2, Sys Tol 0.0003, TUR 666.667 (>= 4.00). 5.001 5700 8.8mA S 2W 5.002 MEMI Use keyboard to enter actual UUT reading in Percent. 5.003 ACC 30.0pctrh 0.0003U 5.004 MEME 5.005 MEMC pctrh 0.25U 8.8mA #! Test Tol 0.25, Sys Tol 0.0003, TUR 833.333 (>= 4.00). 6.001 5700 10.4mA S 2W 6.002 MEMI Use keyboard to enter actual UUT reading in Percent. 6.003 ACC 40.0pctrh 0.0003U 6.004 MEME 6.005 MEMC pctrh 0.3U 10.4mA #! Test Tol 0.3, Sys Tol 0.0003, TUR 1000.000 (>= 4.00). 7.001 5700 12mA S 2W 7.002 MEMI Use keyboard to enter actual UUT reading in Percent. 7.003 ACC 50.0pctrh 0.0003U 7.004 MEME 7.005 MEMC pctrh 0.35U 12mA #! Test Tol 0.35, Sys Tol 0.0003, TUR 1166.667 (>= 4.00). 8.001 5700 13.6mA S 2W 8.002 MEMI Use keyboard to enter actual UUT reading in Percent. 8.003 ACC 60.0pctrh 0.0003U 8.004 MEME 8.005 MEMC pctrh 0.4U 13.6mA #! Test Tol 0.4, Sys Tol 0.0003, TUR 1333.333 (>= 4.00). 9.001 5700 15.2mA S 2W 9.002 MEMI Use keyboard to enter actual UUT reading in Percent. 9.003 ACC 70.0pctrh 0.0003U 9.004 MEME 9.005 MEMC pctrh 0.45U 15.2mA #! Test Tol 0.45, Sys Tol 0.0003, TUR 1500.000 (>= 4.00). 10.001 5700 16.8mA S 2W 10.002 MEMI Use keyboard to enter actual UUT reading in Percent. 10.003 ACC 80.0pctrh 0.0003U 10.004 MEME 10.005 MEMC pctrh 0.5U 16.8mA #! Test Tol 0.5, Sys Tol 0.0003, TUR 1666.667 (>= 4.00). 11.001 5700 18.4mA S 2W 11.002 MEMI Use keyboard to enter actual UUT reading in Percent. 11.003 ACC 90.0pctrh 0.0003U 11.004 MEME 11.005 MEMC pctrh 0.55U 18.4mA #! Test Tol 0.55, Sys Tol 0.0003, TUR 1833.333 (>= 4.00). 12.001 5700 20mA S 2W 12.002 MEMI Use keyboard to enter actual UUT reading in Percent. 12.003 ACC 100.0pctrh 0.0003U 12.004 MEME 12.005 MEMC pctrh 0.6U 20mA #! Test Tol 0.6, Sys Tol 0.0003, TUR 2000.000 (>= 4.00). 13.001 HEAD {} 13.002 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 13.002 DISP {º THIS COMPLETES THE VERIFICATION º} 13.002 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 13.003 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.